User manual

173
Connect up the digital output from your National Instruments M- or X-Series board to your perfusion
system as follows:
1) For the 4-channel pre-flush system (2 valves/line), use these Port1 outputs:
Port1, Bit0 P1.0 PFI 0 Chamber Valve, Ch 1
Port1, Bit1 P1.1 PFI 1 Chamber Valve, Ch 2
Port1, Bit2 P1.2 PFI 2 Chamber Valve, Ch 3
Port1, Bit3 P1.3 PFI 3 Chamber Valve, Ch 4
Port1, Bit4 P1.4 PFI 4 Flush Valve, Ch 1
Port1, Bit5 P1.5 PFI 5 Flush Valve, Ch 2
Port1, Bit6 P1.6 PFI 6 Flush Valve, Ch 3
Port1, Bit7 P1.7 PFI 7 Flush Valve, Ch 4
2) For the 8-channel pre-flush system (2 valves/line), use Port1 and Port2 outputs:
Port1, Bit0 P1.0 PFI 0 Chamber Valve, Ch 1
Port1, Bit1 P1.1 PFI 1 Chamber Valve, Ch 2
Port1, Bit2 P1.2 PFI 2 Chamber Valve, Ch 3
Port1, Bit3 P1.3 PFI 3 Chamber Valve, Ch 4
Port1, Bit4 P1.4 PFI 4 Chamber Valve, Ch 5
Port1, Bit5 P1.5 PFI 5 Chamber Valve, Ch 6
Port1, Bit6 P1.6 PFI 6 Chamber Valve, Ch 7
Port1, Bit7 P1.7 PFI 7 Chamber Valve, Ch 8
Port2, Bit0 P2.0 PFI 8 Flush Valve, Ch 1
Port2, Bit1 P2.1 PFI 9 Flush Valve, Ch 2
Port2, Bit2 P2.2 PFI 10 Flush Valve, Ch 3
Port2, Bit3 P2.3 PFI 11 Flush Valve, Ch 4
Port2, Bit4 P2.4 PFI 12 Flush Valve, Ch 5
Port2, Bit5 P2.5 PFI 13 Flush Valve, Ch 6
Port2, Bit6 P2.6 PFI 14 Flush Valve, Ch 7
Port2, Bit7 P2.7 PFI 15 Flush Valve, Ch 8
3) For the 8-channel standard system (1 valve/line), use these Port1 outputs:
Port1, Bit0 P1.0 PFI 0 Chamber Valve, Ch 1
Port1, Bit1 P1.1 PFI 1 Chamber Valve, Ch 2
Port1, Bit2 P1.2 PFI 2 Chamber Valve, Ch 3
Port1, Bit3 P1.3 PFI 3 Chamber Valve, Ch 4
Port1, Bit4 P1.4 PFI 4 Chamber Valve, Ch 5
Port1, Bit5 P1.5 PFI 5 Chamber Valve, Ch 6
Port1, Bit6 P1.6 PFI 6 Chamber Valve, Ch 7
Port1, Bit7 P1.7 PFI 7 Chamber Valve, Ch 8
4) For the 16-channel standard (system (1 valve/line), use Port1 and Port2 outputs:
Port1, Bit0 P1.0 PFI 0 Chamber Valve, Ch 1
Port1, Bit1 P1.1 PFI 1 Chamber Valve, Ch 2
Port1, Bit2 P1.2 PFI 2 Chamber Valve, Ch 3
Port1, Bit3 P1.3 PFI 3 Chamber Valve, Ch 4
Port1, Bit4 P1.4 PFI 4 Chamber Valve, Ch 5
Port1, Bit5 P1.5 PFI 5 Chamber Valve, Ch 6
Port1, Bit6 P1.6 PFI 6 Chamber Valve, Ch 7
Port1, Bit7 P1.7 PFI 7 Chamber Valve, Ch 8
Port2, Bit0 P2.0 PFI 8 Chamber Valve, Ch 9
Port2, Bit1 P2.1 PFI 9 Chamber Valve, Ch 10
Port2, Bit2 P2.2 PFI 10 Chamber Valve, Ch 11
Port2, Bit3 P2.3 PFI 11 Chamber Valve, Ch 12
Port2, Bit4 P2.4 PFI 12 Chamber Valve, Ch 13
Port2, Bit5 P2.5 PFI 13 Chamber Valve, Ch 14
Port2, Bit6 P2.6 PFI 14 Chamber Valve, Ch 15
Port2, Bit7 P2.7 PFI 15 Chamber Valve, Ch 16
P1.0 and PFI 0 are two different ways National Instruments names Port1, Bit 0 and so forth.