User`s manual
AX2200/AX2300 Instruction Manual Chapter 4 Serial Communications
CM2-AX2000-2001 (Rev.4, 4/2014) 4-17
Registers Variable Data type Units Function
code
Addresses
65100-65101 Serial number unsigned long — 03, 04
30525-30526 Totalizer unsigned long display units* 03, 04 524-525
32037-32042 Totalizer units string — 03, 04 2036-2041
30009-30010 Mass flow float display units* 03, 04 8-9
30007-30008 Volume flow float display units* 03, 04 6-7
30005-30006 Pressure float display units* 03, 04 4-5
30001-30002 Temperature float display units* 03, 04 0-1
30029-30030 Velocity float ft/sec 03, 04 28-29
30015-30016 Density float display units* 03, 04 14-15
30013-30014 Viscosity float cP 03, 04 12-13
30031-30032 Reynolds number float — 03, 04 30-31
30025-30026 Vortex frequency float Hz 03, 04 24-25
34532 Gain char — 03, 04 4531
30085-30086 Vortex amplitude float Vrms 03, 04 84-85
30027-30028 Filter setting float Hz 03, 04 26-27
Table 4-2.Register Definitions
The following registers are available with the energy meter firmware:
Registers Variable Data type Units Function
code
Addresses
30527-30528 Totalizer #2 unsigned long display units* 03, 04 526-527
32043-32048 Totalizer #2 units string — 03, 04 2042-2047
30003-30004 Temperature #2 float display units* 03, 04 2-3
30011-30012 Energy flow float display units* 03, 04 10-11
The following registers contain the display units strings:
Registers Variable Data type Units Function code Addresses
32007-32012 Volume flow units string — 03, 04 2006-2011
32001-32006 Mass flow units string — 03, 04 2000-2005
32025-32030 Temperature units string — 03, 04 2024-2029
32019-32024 Pressure units string
—
03, 04 2018-2023
32031-32036 Density units string — 03, 04 2030-2035
32013-32017 Energy flow units string — 03, 04 2012-2017
Function codes 03 (read holding registers) and 04 (read input registers) are the
only codes supported for reading these registers, and function codes for writing
holding registers are not implemented. We recommend that the floating point and
long integer registers be read in a single operation with the number of registers
being a multiple of two. If these data are read in two separate operations, each