User`s manual
Chapter 3 Operation AX2200/AX2300 Instruction Manual
3-8 CM2-AX2000-2001 (Rev. 4, 4/2014)
Display Menu
Password
ENTER
Display
Menu
Cycle Time (sec)
0
ENTER
Run Mode
Number of Digits
2
MF Vf Te Pr De T
Y or N
A1 A2 A3 Fl Dt E
Y or N
If Cycle Time is set to zero, manual advance is required
Used to set the number of digits displayed after the decimal point
MF = Mass Flow
Vf = Volume Flow
Te = Temperature
Pr = Pressure
De = Density
T = Total
A1 = Alarm 1 Status
A2 = Alarm 2 Status
A3 = Alarm 3 Status
Fl = Fluid
Dt = Density
For each parameter:
Select Yes to view parameter in Run Mode
Select No to hide parameter in Run Mode
Display TC (sec)
1
TC = Display Time constant, used to smooth display
*
E = Energy
* Energy EMS Meters Only
Use
keys to access menus
Use the Display Menu to set the cycle time for automatic screen sequencing used in the Run
Mode, change the precision of displayed values, smooth the values or enable or disable each item
displayed in the Run Mode screens.
Example for Changing a Run Mode Display Item
The following shows how to remove the temperature screen from the Run Mode screens. Note: all outputs
are disabled while using the Setup Menus.
1. Use keys to move to the Display Menu.
2. Press key until Mf Vf Pr Te De T appears.
3. Press
ENTER to select. Press key until the cursor is positioned below Te.
4. Press key until N appears. Press
ENTER to select.
5. Press EXIT and then
ENTER to save changes and return to the Run Mode.