User`s manual

Appendix A Specifications AX2200/AX2300 Instruction Manual
A-4 CM2-AX2000-2001 (Rev. 4, 4/2014)
AX2300 Pressure Ratings
Probe Seal Process
Connection
Material Rating Ordering
Code
Compression
Fitting
2-inch MNPT
316L SS
ANSI 600 lb
CNPT
2-inch 150 lb flange,
DN50 PN16
316L SS
ANSI 150
lb, PN16
C150,
C16
2-inch 300 lb flange,
DN50 PN40
316L SS
ANSI 300
lb, PN40
C300,
C40
2-inch 600 lb flange,
DN50 PN64
316L SS
ANSI 600
lb, PN64
C600,
C64
Packing Gland 2-inch MNPT 316L SS 50 psig PNPT
2-inch 150 lb flange,
DN50 PN16
316L SS 50 psig
P150,
P16
2-inch 300 lb flange,
DN50 PN40
316L SS 50 psig
P300,
P40
Packing Gland with
Removable Retractor
2-inch MNPT
316L SS
ANSI 300 lb
PM, RR
2-inch 150 lb flange,
DN50, PN16
316L SS ANSI 150 lb
P150,
P16,RR
2-inch 300 lb flange 316L SS ANSI 300 lb
P300,
P40, RR
Packing Gland with
Permanent Retractor
2-inch MNPT
316L SS
ANSI 600 lb
PNPTR
2-inch 150 lb flange,
DN50 PN16
316L SS ANSI 150 lb
P150R,
P16R
2-inch 300 lb flange,
DN50, PN40
316L SS ANSI 300 lb
P300R,
P40R
2-inch 600 lb flange,
DN50 PN64
316L SS ANSI 600 lb
P600R,
P64R
Pressure Transducer Ranges
Pressure Sensor Ranges
(1)
, psia (bara)
Full Scale Operating Pres-
sure
Maximum Over-Range Pressure
psia (bara) psia (bara)
30 2 60 4
100 7 200 14
300 20 600 40
500 35 1000 70
1500 100 2500 175
Note: (1) To maximize accuracy, specify the lowest full scale operat-
ing pressure range for the application. To avoid damage, the
flow meter must never be subjected to pressure above the
over-range pressure shown above.
Power Requirements 12 to 36 VDC, 25 mA, 1 W max., Loop Powered Volumetric or Mass
12 to 36 VDC, 300 mA, 9 W max. Multiparameter Mass options
100 to 240 VAC, 50/60 Hz, 5 W max. Multiparameter Mass
options
Class I Equipment (Grounded Type)
Installation (Over-voltage) Category II for transient over-voltages
AC & DC Mains supply voltage fluctuations are not to exceed
+/-10% of the rated supply voltage range.