User`s manual

AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair
CM2-AX2000-2001 (Rev. 4, 4/2014) 5-15
Pressure Sensor Replacement (AX2200 Models Only)
1. For local mounted electronics, remove the electronics stack as
previously described. For remote mount electronics, remove
all wires and sensor connectors from the remote feed through
board in the junction box at the meter.
2. Loosen the three set screws at the center of the adapter between
the meter and the enclosure.
3. Remove the top half of the adapter to expose the pressure
transducer.
4. Remove the transducer and replace it with the new one using
appropriate thread sealant.
5. Reassemble in reverse order.
Returning Equipment to the Factory
Before returning any AX2000Series flow meter to the factory,
you must request a Return Material Authorization (RMA)
number. To obtain an RMA number and the correct shipping
address, contact Customer Service at:
888-262-4639 or 602-216-8199 in the USA,
When contacting Customer Service, be sure to have the meter
serial number and model code.
Please see the Meter Troubleshooting Checklist for additional
items which may help with problem isolation. When requesting
further troubleshooting guidance, please record the values on the
checklist at no flow and during flow if possible.