User`s manual
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair
CM2-AX2000-2001 (Rev. 4, 4/2014) 5-13
Figure 5-4.Temperature Sensor Connector
4. Consult factory with findings
Symptom: Meter Displays Pressure Fault
1. For remote mounted electronics, carefully check all the wiring
connections in the remote mount junction box. There are 18
connections that must be correct, verify each color (black and
red), shield, and wire number.
2. Using ESD precautions and hazardous area precautions,
remove the electronics enclosure window cover. Disconnect
the pressure sensor from the electronics stack or the remote
feed through board. Measure the resistance across the outside
pins of the pressure sensor connector, then across the inside
pins. Both readings should be approximately 4000 ohms.
Figure 5-5.Pressure Sensor Connector
3. Go to the first level of the hidden diagnostics and record the
Pe(V) and Pv(V) values and consult the factory with findings.
OUTSIDE
OUTSIDE
OUTSIDE
OUTSIDE
INSIDE