AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair
CM2-AX2000-2001 (Rev. 4, 4/2014) 5-7
Init Displ. (sec) = Enter a value in seconds to initialize the
display every xxx seconds. Enter a value of 0 to disable
initializing the display.
Analog Output Calibration
To check the 4–20 mA circuit, connect a DVM in series with the
output loop. Select zero or full scale (from the second level of the
hidden diagnostics) and then actuate the enter key twice. This
action will cause the meter to output its 4 mA or 20 mA condition.
If the DVM indicates a current greater than ± 0.006 mA from 4 or
20, adjust the setting up or down until the output is calibrated.
Note: these settings are not for adjusting the output zero and span
to match a flow range, that function is located in the Output Menu.