User`s manual
Chapter 5 Troubleshooting & Repair AX2200/AX2300 Instruction Manual
5-6 CM2-AX2000-2001 (Rev. 4, 4/2014)
Factory Defaults = Reset factory defaults. If you change this
to Yes and press Enter, all the factory configuration is lost and
you must reconfigure the entire program. Consult the factory
before performing this process, it is required only in very rare
cases.
Meter Type = Insertion (AX2300) or Inline (AX2200) meter.
Config Code = Factory use only.
Test Pulse Out = Force totalizer pulse. Set to Yes and press
enter to send one pulse. Very useful to test totalizer counting
equipment.
Test Scaled Freq = Enter a frequency value in order to test the
scaled frequency output. Return to 0 to stop the test.
Output Type = Factory use only.
Calibration Mode = Factory use only.
A2D Ref. Resistor = Factory use only.
Pressure Cal Current = Calibration value for the electronics
and pressure transducer combination. Consult Factory for
value.
Pressure 9Cs = Nine pressure coefficients unique to the
pressure transducer. Use the RIGHT ARROW to access all
nine coefficients.
o Press. Max psi = Based on installed sensor.
o Press. Min psi = 0 psia
Press. Min psi = 0 psiaRTD1. Press the RIGHT ARROW to
access:
o Ro = RTD resistance at 0°C (1000 ohms).
o A = RTD coefficient A (.0039083).
o B = RTD coefficient B (-5.775e-07).
o RTD1 Max Deg. F = 500
o RTD1 Min Deg. F = -330
RTD2 = Second RTD configuration, for special applications
only.
Correction Pairs
o ft3/sec (1 through 10)
o %Dev. (1 through 10)
Roughness = Factory use only.
Force Recal? = Factory use only.
Min. Delta H – Energy EMS meters only. Sets the deadband
for totalization to begin. Must be greater than this number (1
default) to initiate the totalizer.