User`s manual
AX2200/AX2300 Instruction Manual Chapter 5 Troubleshooting & Repair
CM2-AX2000-2001 (Rev. 4, 4/2014) 5-1
Chapter 5 Troubleshooting and Repair
Hidden Diagnostics Menus
The menus shown on the following page can be accessed using the password
16363, then moving to the display that reads “Diagnostics Menu” and pressing
ENTER (rather than one of the arrow keys).
Use the right arrow key to move to the second level. Press EXIT to move from
the second level back to the first, press EXIT while in the first level to return to
the setup menus.
Caution: password 16363 will allow full access to the configuration and should
be used carefully to avoid changes that can adversely alter the function of the
meter.
Each of the menus on the following page will first be defined followed by
specific troubleshooting steps.
Warning!
Before attempting any flow
meter repair, verify that the
line is not pressurized.
Always remove main power
before disassembling any
part of the mass flow meter.