Installation guide

Graphical User Interface
Advanced Dicing Technologies Ltd.
2.2-32
Overflow Sensor
In the Setup & Diagnostics tree, under Saw, the Overflow sensor is
displayed. This sensor is intended to prevent the damage to the System due
to the drain pan overflow, which may be caused by clogged drain outlet
If the drain pan becomes overflown, the indicator (see Figure 2-7) changes
its color from gray to red, and an error message appears (see Figure 2-8).
Once an overflow is indicated the di-water supply to the cooling block and
to the Non-Contact Height Device becomes disabled. During the time that
the overflow sensor is on, the operator cannot operate the water. For more
details about the Overflow sensor, refer to the 7100 Series Maintenance
Manual.
For a detailed description of the setup parameters and diagnostic tests for
each station, refer to the ADT Model 7100 Semi-Automatic Dicing System
Maintenance Manual.
Figure 2-7: Overflow Sensor Indication
Figure 2-8: Water Overflow Message