Installation guide
Tilted Spindle
Advanced Dicing Technologies Ltd.
9.2-10
9.2.4.4.4 Sample Blade Calibration
To Perform Sample Blade Calibration
1Open the Blade screen.
2 Enter the current Blade diameter as the Sample Blade diameter.
3Open the Height screen.
4 Perform Sample Blade Calibration.
5 Click Save.
9.2.4.4.5 Chuck to Non-Chuck Height
To Perform Chuck to Non-Chuck Height Procedure
1 In the Height screen, click Chuck to Non-Chuck Height.
2 Click Save.
9.2.4.4.6 Teaching the New Y-Offset
To Teach the New Y-Offset
1 Perform a single cut.
2 Click the Y-Offset calibration icon.
Note: Before performing Sample Blade Calibration, update the Z Calibration
start position according to the new Blade height.
Note: Several Chuck to Non-Chuck measurements can be carried out, to
increase accuracy, as described in
section 7.3.3.1.