Installation guide

ADT Model 7100 Semi-Automatic Dicing System Operations Manual
Autoloader
P/N 97100-9002-000-14 Ver 06/05
9.1-17
The system notifies the user by
Displaying an error message
Flashing the yellow light on the Light Tower
Indicating the fault location by illuminating the appropriate virtual
LEDs on the Model 7100 window.
The system does not resume operation until the problem is corrected,
following the steps outlined below.
To Recover From a System Failure:
1 Remove the frame that was being processed at the time the failure
occurred. The frame must be removed regardless of its location in the
system.
2 In the error message box, click OK. The Model 7100 initiates.
3 Click Run.
4 If the problem persists, use the Troubleshooting Flow Diagram
given in section 14.9 of the Maintenance Manual to isolate the problem
and correct it.
Note: In the case of a frame improperly positioned in the Cassette, the frame
may be repositioned without removing it from the system.
Note: The system resumes operation with the next available frame.