Installation guide
ADT Model 7100 Semi-Automatic Dicing System Operations Manual
Workpiece Alignment
P/N 97100-9002-000-14 Ver 06/05
6.1-7
8 Define a Cut Position (see section 6.1.1.2). At the end, click Finish.
The following dialog box is displayed:
9 Click Auto.
10 Click Next. The Camera moves to the top of the Workpiece.
11 Select a Model to be taught by moving it to the center of the FOV using
the X/Y-Axis Controls.
12 Toggle the center box of the Guide Control until T appears, then use
the Guide Control to size the Teach Area.
13 Adjust the Focus and Illumination for the best view of the model.
14 Select the confidence level for the model.
15 Click Teach. If the Model selected is successfully taught to the
System, the FOV will flash the message Model Taught, along with a
score (based on a scale of 0-100). Click Find to test whether the
System can correctly identify the Model.
16 Click Next. When updating an existing index, the System moves one
Index along the Y-Axis, based on the value currently defined for the
Index parameter in the Recipe and searches for the taught Model.
When teaching a new index, the System moves one millimeter along
the Y-Axis.
17 Use the X/Y-Axis Controls to adjust the position of the Reticle along
the Y-Axis until the Model is approximately centered inside the Search
Area.
18 Click Finish. The System automatically attempts to find the Model at
each of five consecutive Index jumps. After each jump, the System
Figure 6-7: Teach Index Process Dialog Box
Note: Click Auto if there are clearly visible models on the substrate. Clicking
Manual enables the User to teach the Index manually as described in
section
6.1.2.1.1.
Note: In order to verify the adjustment before completing the procedure,
another manual Index move can be made by clicking Next.