User guide
DuoSpray™ Ion Source User Reference
System User Guide
D5033331 A
AB SCIEX TripleTOF
®
5600/5600+ Instruments
99 of 116
Remove the Corona Discharge Needle
The corona discharge needle tip may become so corroded that it must be cut off from the corona
discharge needle. If this occurs, replace the corona discharge needle.
1. Remove the ion source and probe from the mass spectrometer.Refer to Remove the
Ion Source on page 96.
2. Rotate the ion source so that the open side is toward you.
3. Press down on the corona discharge needle adjustment knob on the top of the
tower. The corona discharge needle extends.
4. Holding the corona discharge needle tip between the thumb and forefinger of one
hand and the corona discharge needle with the other hand, rotate the corona
discharge needle tip counter-clockwise to loosen and gently remove the tip.
Figure 7-5 Electrode tip extension adjustment
Item Description
1 Probe
2 Electrode
WARNING! Electrical Shock Hazard: Remove the ion source from the mass
spectrometer before starting any maintenance procedures.
WARNING! Piercing Hazard: The tip of the needle is extremely sharp. Take care
to handle it safely.
0.5 - 1 mm
1
2