User guide
DuoSpray™ Ion Source User Reference
AB SCIEX TripleTOF
®
5600/5600+ Instruments
86 of 116
System User Guide
D5033331 A
Gas and Electrical Connections
Gas and high-voltage electrical connections enter through the front plate of the interface and
connect internally through the ion source housing. When the ion source is installed on the mass
spectrometer, all of the electrical and gas connections are made.
Ion Source Latch
A latch disables the high-voltage power supply for the mass spectrometer and the source
exhaust system if:
• The ion source housing is not installed or is improperly installed.
• A probe is not installed.
• The mass spectrometer senses a gas fault.
Source Exhaust System
A passive pressure exhaust system removes ion source gases through a drain port without
introducing chemical noise. The drain port connects through a drain chamber and a source
exhaust pump to a drain bottle, and from there to a customer-supplied exhaust ventilation
system. For more information on the requirements for the source exhaust system, refer to the
Site Planning Guide for the mass spectrometer.
Figure 7-3 Parts of the APCI probe
Item Description
1 Electrode adjustment nut (black collar) that adjusts the extension of electrode tip
2 Bronze retaining ring that fastens the probe to the probe tower on the ion source
housing
3 Electrode tip through which samples are sprayed into the sample inlet area of the ion
source
WARNING! Toxic Chemical Hazard: Be sure to use the source exhaust system to
safely remove sample vapor exhaust from the laboratory environment. For
requirements for the source exhaust system, refer to the AB SCIEX TripleTOF
®
5600/5600+ Instruments Site Planning Guide.
2
1
3