User guide
Analyzing and Processing Data
AB SCIEX TripleTOF
®
5600/5600+ Instruments
72 of 116
System User Guide
D5033331 A
2. Click Explore > Extract Ions > Use Dialog.
Figure 6-2 Extract Ions dialog
3. Type the values for each XIC to be created. If a stop value is not typed, then the
range is defined by the start value.
• In the Start field, type the start value (lower value) for the mass range to be
extracted.
• In the Stop field, type the stop value (higher value) for the mass range to be
extracted.
4. Click OK.
An XIC of the selection opens below the chromatogram pane. The experiment
information at the top of the pane includes the masses and the maximum intensity in
counts per second.
Generate BPCs
BPCs can be generated only from single period, single experiment data. For more information
about using the available icons, refer to Table 6-8.
1. Select an area within a TIC.
The selection is highlighted in blue.