User guide
Analyzing and Processing Data
System User Guide
D5033331 A
AB SCIEX TripleTOF
®
5600/5600+ Instruments
71 of 116
To generate an XIC using the maximum peak
1. In the Navigation Bar, under Explore, double-click Open Data File.
The Select Sample dialog opens.
2. In the Data Files list, select a file containing spectra.
3. Click OK.
4. Select a range.
The selection is highlighted in blue.
5. Click Explore > Extract Ions > Use Maximum.
An XIC of the maximum peak specified selection opens below the spectrum pane.
The experiment information at the top of the pane contains the mass range and the
maximum intensity in counts per second.
To generate an XIC using base peak masses
1. In a BPC, select the peak from which to extract ions.
The selection is highlighted in blue.
2. Click Explore > Extract Ions > Use Base Peak Masses.
An XIC of the specified selection opens below the spectrum pane. The experiment
information at the top of the pane shows the mass range and the maximum intensity
in counts per second.
To extract ions by selecting masses
1. Open a spectrum or chromatogram.