User guide
Analyzing and Processing Data
AB SCIEX TripleTOF
®
5600/5600+ Instruments
68 of 116
System User Guide
D5033331 A
3. Click Explore > Show > Show ADC Data.
Figure 6-1 Select ADC Channel dialog
4. In the Channel list, select a channel.
5. Click OK.
The ADC data opens in a new pane below the active pane.
Show Basic Quantitative Data
1. With a data file open, click Explore > Show > Show List Data.
2. In the Peak List tab, right-click and select Show Peaks in Graph.
Peaks appear in two colors.
3. To change the peak finding algorithm settings, right-click and then select either
Analyst Classic Parameters or Intelliquan Parameters, which ever is active.
4. (Optional) To remove the colored peaks, right-click in the Peak List tab and then
clear Show Peaks in Graph.
Chromatograms
Table 6-4 Chromatograms
Types of chromatograms Purpose
TIC (Total Ion Chromatogram) A chromatographic display generated by plotting the intensity
of all ions in a scan against time or scan number.
When a data file is opened, it is preset to appear as a TIC. If
the experiment contains only one scan, it is shown as a
spectrum. For more information about using the available
icons, refer to Table 6-8.
If the MCA check box is selected during acquisition of the data
file, then the data file opens to the mass spectrum. If the MCA
check box is not selected, then the data file opens with the TIC
XIC (Extracted Ion
Chromatogram)
An ion chromatogram created by taking intensity values at a
single, discrete mass value, or a mass range, from a series of
mass spectral scans. It indicates the behavior of a given mass,
or mass range, as a function of time.