User guide
Instrument Tuning and Calibrating
AB SCIEX TripleTOF
®
5600/5600+ Instruments
40 of 116
System User Guide
D5033331 A
4. Select a tuning solution. Make sure that the tuning solution matches the reference
table.
5. The Verify Performance Only check box is preselected. Click Next.
For this example, leave this option selected. If the report indicates that the
instrument needs tuning, then run Instrument Optimization again and select one or
more scan modes to optimize.
6. Make sure that the ion source and syringe parameters are suitable.
7. Click GO.
The Verifying Performance screen appears. After the process has completed, the
Results Summary appears showing the resolution and intensity for each scan mode.
Note: Users can also use the CDS to inject the solution. Make sure the
tuning solution matches the configuration in the reference table. Set the
appropriate flow rate and then click CDS Inject.
Note: Ensure that the correct Calibrant Valve Position is selected in the
Reference Table Editor for the chosen reference table. CDS can select from
up to four different positions, A to D.