3. Turn the side screws as shown in Figure 1-7 to secure the syringe.
Figure 1-7 Syringe pump
4. In the Analyst TF software, on the Navigation bar, double-click Manual Tuning.
5. Click Start Syringe.
6. To stop the syringe pump, click Stop Syringe.
Configure the Integrated Syringe Pump
1. In the Navigation bar, under Acquire, double-click Build Acquisition Methods.
2. In the Acquisition method pane, click the Syringe Pump icon.
The Syringe Pump method properties tab opens in the Acquisition Method Editor
pane.
3. In the Syringe Diameter (mm) field, type the syringe diameter.
4. In the Flow Rate field, type the flow rate.
5. In the Unit list, select the units of flow.
6. Save the file.
2 Post. Adjust the height to prevent the syringe plunger from hitting the
syringe during sample infusion.
3 Post lock screw. Tighten the screw after you have adjusted the height of
the post.