Clean the Curtain Plate
Caution: Potential Instrument Damage: Do not rest the curtain plate on the orifice. Make
sure that the conical side faces up.
1. Remove the curtain plate and then place it, conical side up, on a clean, stable
surface.
Figure 8-4 Interface with curtain plate removed
2. Using wipes and water, clean both sides of the curtain plate.
3. Repeat step 2 using the cleaning solution.
4. Using a dampened wipe or small poly swab, clean the aperture.
5. Wait until the curtain plate is dry.
Figure 8-3 Source drain on the vacuum interface
Item Description
1 Source drain