5. Install the ion source. Refer to Install the Ion Source on page 88.
6. Resume the sample flow.
Troubleshooting
Table 7-4 Troubleshooting
Symptom Possible cause Solution
The Analyst
®
TF software
reports that the mass
spectrometer is in Fault state.
(The mass spectrometer icon on
the Analyst TF software status
bar is red.)
• The probe is not installed.
• The probe is not connected
securely.
• Install the probe. Refer to
Install the Probe in the Ion
Source Housing on
page 87.
• Remove and replace the
probe. Tighten the probe
connection bronze ring
securely. Refer to Remove
the Probe on page 96 and
Install the Probe in the Ion
Source Housing on
page 87.
The heater does not work. The F3 fuse is blown. Contact your FSE.
The spray does not appear to be
uniform.
The electrode is blocked. Clean or replace the
electrode. Refer to Clean the
Electrode Tube on page 97.
Sensitivity is poor. Excess high voltage induces
fragmentation before the ions
enter the mass filters.
Optimize ISVF or DP. Refer
to Optimize the
TurboIonSpray Probe on
page 90 or Optimize the
APCI Probe on page 92.
During testing, the ion source
fails to meet specifications.
The mass spectrometer has
not passed the installation
tests.
The test solution was not
prepared correctly.
Perform installation tests on
the mass spectrometer with
the default source.
Confirm that the test
solutions were prepared
correctly.
If the problem cannot be
resolved, contact an FSE.
Background noise is high. • Temperature (TEM) is too
high.
• Gas 2 flow rate (GS2) is too
high.
• Optimize TEM.
• Optimize GS2.