User`s manual

9-47
IM 701610-01E
Waveform Analysis
9
Setting the Logic
20. Turn the jog shuttle to move the cursor to Logic.
21. Press SELECT to select AND or OR.
Setting the Sequence
22. Turn the jog shuttle to move the cursor to Sequence.
23. Press SELECT to select Single or Continue.
Setting the ACQ Count
24. Turn the jog shuttle to move the cursor to ACQ Count.
25. Pressing SELECT a menu used to set the number of waveform acquisitions.
26. Turn the jog shuttle to set the number of waveform acquisitions and press
SELECT.
Selecting the Action to Take when the Condition is not Met
27. Turn the jog shuttle to move the cursor to Action.
28. Press SELECT to turn the action ON or OFF.
Setting the Determination Range
29. Press the T-Range1/T-Range2 soft key to set the jog shuttle control to T-Range1.
30. Turn the jog shuttle to set the start point of the determination range.
31. In a similar fashion, set the end point of the determination range with T-Range2.
Executing the GO/NO-GO Determination
32. Press the Exec soft key to execute the GO/NO-GO determination.
When the determination terminates, the waveform acquisition is automatically
stopped.
To forcibly stop the operation, press START/STOP or the Abort soft key to stop
the acquisition.
Explanation
Waveform Parameters that can be Determined using GO/NO-GO Operation: Item
This function is applicable to the all of waveform parameters (27 items) in section 9.2.
Up to four types of parameters can be determined simultaneously.
Mode
OFF: Does not perform GO/NO-GO determination
ON: Performs GO/NO-GO determination
Upper and Lower Limits of Parameters: Upper/Lower
The upper and lower limits depends on the parameters. They can be set in the range
– 9.9999E+30 to 9.9999E+30.
Setting the Condition
IN: When the value is inside the upper and lower limits.
OUT: When the value is outside the upper or lower limits.
9.9 GO/NO-GO Determination Using the Measurement of Waveform Parameters