Spec Sheet

Church Road,
Weston on the Green,
Oxfordshire, OX25 3QP, U.K.
Tel +44 (0)1869 351 155
email: sales@axicon.com
www.axicon.com
Axicon 6015 Barcode Verier
SOFTWARE SPECIFICATIONS
Symbologies Veried: GS1 SYMBOLOGIES: EAN-8, EAN-13 (with or without addons), ITF-14/Case Code, GS1-DataBar (all symbologies), GS1-
128, UPC-A, UPC-E (with or without addons) (subject to length of barcode).
OTHER SYMBOLOGIES: Code 39, Code 93, Code 128, Codabar, ITF, MSI Plessey, Pharmacode (Laetus: optional extra).
Application Standards: AIAG, Belgian Pharmacode (MSI), Belgian Pharmacode (Unique), CIP39, Coupon Codes (UK, USA, Euro),
Code Vignette, GS1 AI Check, HIBC, ISBN/ISSN, Italian Pharmacode (IMH), LPPR, M&S, PZN, SISAC, Variable Measure
Codes (Branded, Instore, Australian).
Analysis Performed
on Relevant Symbologies: Full ISO/ANSI Parameter Analysis (Rmin, Rmax, Global Threshold, Symbol Contrast, Min Edge Contrast, Modulation,
Defects, Decodability, Decode), Decodability per Symbol Character, Average Bar Gain, Check Digit Validation, Data
Length, Nominal Bar Width Analysis (X Dimension), Print Contrast Measurement (PCS), Symol Structure including
Parity, Subsets & Start/Stop Characters, Quiet Zone Validation, Wide to Narrow Ratio Display and Validation.
Diagnostics Windows: Parameter Detail Results, Dimensional Analysis, Pass/Fail, Scan Reectance Prole, Static Scan
Reetance, Summary, Traditional Results, Application Identier Data Content Checker (for GS1-128,
GS1-DataBar), Product-Look-Up.
Logging: Scan Saving and Loading - stores complete details of scan le data. Allows for remote diagnostics and
also for increased traceability.
Calibration Logging: A record of date and time to show when the hardware has been calibrated is kept - vital for ISO 9000 records.
Congurable Options: Audible Warnings, Calibration Expiry Warning, Changeable Pass / Fail and Warning Colours, Optional Check Digit
Validation (Code 39, Codabar, ITF), File Saving, Multi-Scan Averaging, Pass Grade, Print Options, Window Sizes and
Positions.
Error Messages: Check Digit Error, Data Content Error, Incorrect Wide to Narrow Ratio, Print Quality Failure, Product
Look-Up Error, Quiet Zone Error, Reference Decode Algorithm Error, Structure Error, X Dimension Range Checking.
Plugins Available: Code 39 full ASCII, CSV File Saving, Gen Spec, Job Reference, Product Look-up, Text Look-up, Time Zone,
UPC, User Data, User Name, (see also Application Standards).
Additional Software: ScanDB – Reporting Tool transferring data from saved scans into a .csv le.
RepGen – Congurable report generation tool.
Pharmacode (Laetus) – Decode and Analysis of Laetus Pharmacode (chargeable optional extras).
HARDWARE SPECIFICATIONS
Dimensions: 165 x 75 (50) x 128 mm, (6.4” x 2.9” (2”) x 5”), Scan Width – 68mm / 2.6” including quiet zones
Weight: 230g (8.11oz)
Construction: ABS plastic
Interface: USB
Power Consumption: Current draw 110mA – supplied from USB host.
Nominal Measuring: 3mil, 5mil, 6mil, 10mil and 20mil.
Aperture: The software automatically identies the symbology/nominal bar size (X dimension) and adjusts the aperture size in
order to comply with the symbology specications and International Standards.
Wavelength of Light: 660 nanometers
GENERAL INFORMATION
Print of Analysis: A printed report can be produced on any connected windows printer. The report will be automatically resized to suit
the output media. Printouts can be produced for the main verication information (including: Decoded Number,
Time & Date, Number of Scans, Code Type, Average ISO/ANSI Grade, Average Bar Gain, Magnication, Check
Character, Quiet Zone & Individual Parameters) Scan Reectance Proles, Dimensional Analysis, also for GS1-128 &
GS1-DataBar Data Content information and Plugin Results.
GS1 Verication Template supported.
Customised reports available via RepGen and ScanDB utilities.
Standard Accessories: Transit Case, Calibration Sheet, Start-up Guide, Software CD, Compliance Certicates.
System Requirements: PC running Windows 95 (SP2) or later, Mac OSX 10.4 or later (but not “Lion”)
Environmental: Operating Temperature: 10° to 40°C
Storage Temperature: -20° to 70°C
Relative Humidity: 25% to 80% @ 40°C non-condensing
Regularity approvals: CE Certied, ROSH Compliant
Meets ISO/IEC-15416 and 15426-1 standards
Has all the features necessary to meet the requirements of 21 CFR part 11.
Warranty: 2 year parts and labour - return to bench.
All trademarks are acknowledged as belonging to their respective companies.
Due to Axicon’s continuing product improvement programs, specications and features herein are subject to change without notice.
Please note that not all features are available on Mac OS
Oct 2012