Specifications

LabMaster 9 Zi-A
130
922157-00 Rev A
GLITCH
Glitch trigger is a simpler form of Width trigger. Use Glitch trigger when
you want to define a fixed pulse-width time or time range only. Glitch
trigger makes no provision for voltage levels or ranges.
DROPOUT
Used primarily in single-shot applications, and usually with a pre-trigger
delay, Dropout trigger can detect lost signals. The trigger is generated at
the end of the timeout period following the last trigger source transition.
RUNT
The Runt trigger occurs when a pulse crosses a first threshold line, but fails
to cross a second threshold line before re-crossing the first. Other defining
conditions for this trigger are the edge (triggers on the slope opposite to
that selected) and runt width.
SLEW RATE
SlewRate trigger activates a trigger when the rising or falling edge of a
pulse crosses two threshold levels: an upper level and a lower level. The
pulse edge must cross the thresholds faster or slower than a selected
period of time.