18
Immunity
Standard Test Description Test Level/Limits
IEC 61000-4-3 Radio Frequency
Electromagnetic Field
Amplitude Modulated
80 MHz to 2.7 GHz
3V/m
80% AM 1kHz
Spot frequency tests 385 MHz (18 Hz Pulse mod) 27 V/m, 450
MHz (1 kHz sine FM +/-5 kHz dec) 28 V/m,
710, 745 & 780 MHz (217 Hz Pulse mod)
9 V/m, 810, 870 & 930 MHz (18 Hz Pulse
mod) 28 V/m, 1.720, 1.845, 1.970 & 2450
GHz (217 Hz Pulse mod) 28 V/m, 5.240,
5.500 & 5.785 GHz (217 Hz Pulse mod) 9
V/m
IEC 61000-4-4 Electrical Fast Transients ±2.0kV AC Mains
IEC 61000-4-5 Surge 1kV Line to Line (D M)
2kV Line to PE (CM)
IEC 61000-4-6 Radio Frequency Common
Mode
0.150 to 90 MHz
3Vrms
(6 Vrms in ISM bands)
80% AM 1kHz
IEC 61000-4-8 Power Frequency Magnetic
Field
30A/m
IEC 61000-
4-11
Voltage Dips
and Interruptions
0% of Vnom 0.5 cycles
(Phase angles 0, 45, 90, 135, 180, 225,
270 & 315)
0% of Vnom
1.0 cycle
70% Vnom
25 cycles
0% of Vnom
250 cycles