User manual

23
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Philips Avent breast pump uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Philips Avent breast pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
uctuations/ icker
emissions IEC
61000-3-3
Complies
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specied
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic environment
– guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact± 8
kV air
± 6 kV contact± 8
kV air
Floors should be wood, concrete
or ceramic tile. If oors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
± 2 kV for power
supply lines
± 2 kV for power
supply lines
Mains power quality should be
that of a typical residential or
hospital environment.
SurgeIEC 61000-4-5 ± 1 kV line(s) to
line(s)
± 1 kV line(s) to
line(s)
Mains power quality should be
that of a typical residential or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5 % UT (>95 %
dip in UT) for 0,5
cycle40 % UT (60
% dip in UT) for 5
cycles70 % UT (30
% dip in UT) for
25 cycles<5 % UT
(>95 % dip in UT)
for 5 s
<5 % UT (>95 %
dip in UT) for 0,5
cycle70 % UT (30
% dip in UT) for
25 cycles<5 % UT
(>95 % dip in UT)
for 5 s
Mains power quality should be
that of a typical residential or
hospital environment. If the user
of the Philips Avent breast pump
requires continued operation
during power mains interruptions,
it is recommended that the Philips
Avent breast pump be powered
from an uninterruptible power
supply or a battery.Interruptible
electric supply can cause the
breast pump to turn o. This is
acceptable as this will not lead to
unacceptable risks.
Power
frequency(50/60 Hz)
magnetic eldIEC
61000-4-8
3 A/m 3 A/m Power frequency magnetic
elds should be at levels
characteristic of a typical location
in a typical residential or hospital
environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
English