User's Manual
Figure 187: Example: eESPA module acts as output program
An authorized of an entry typically found in this field is as follows: 2:3^3. This indicates that the
first 3 characters of the display message, a ^ and the values of data identifier 3 is equal to the
alarm description. The value NUR^1, NUR^2, SAN^1, and so on, must be configured in the
eKERNEL_ALARM table.)
eESPA_Ala_descr_default_str
This field is used to provide a default alarm description, in the event that no value can be
retrieved from the available datastream. In this case, the same value is used for all alarms
generated through this eESPA interface instance. This alarm description refers to the
definitions of eKERNEL_ALARM table.
This parameter is only relevant if eESPA acts as an input program (so receives external data
from the espa infrastructure.
An authorized of an entry typically found in this field is as follows: ESPA
eESPA_Remove_after_str
This parameter accepts values *SENT, *RESET, or *CALC.
This parameter is only relevant if eESPA acts as an input program (so it receives external data
from the espa infrastructure).
In most cases the eESPA interfaces is used to capture alarms and received data contains
alarm information (acts as input program). In this situation messages are transmitted to
eESPA parameters
DECT Messenger Installation and Commissioning — Book 2 March 2012 233