Specifications
J326XP-IM-5AB
4-6
Troubleshooting
Accessing the Rotor in Case of Power Failure
1
Turn the power switch to off (O) and unplug the power cord from the main source.
2
Insert a 4-mm or
5
/32-in. Allen wrench straight through one of the holes (either one) and turn
the wrench to the left (counterclockwise) about one-half turn.
• The latch disengages.
3
Repeat step 2, inserting the Allen wrench through the other hole.
• After the second latch disengages, the front panel will fall forward about an inch from
the top.
• The bottom of the front panel is held in place by three grooved brackets, attached to the
front panel, that seat over a metal retaining strip on the centrifuge bottom panel.
4
Lift the front panel up and off the centrifuge; set it aside.
• You will see an inner front panel that extends about halfway down from the centrifuge top
(see Figure 4.2).
• Below this inner panel is a black manual door release interlock lever (see Figure 4.2).
Figure 4.2 Manual Door Release
1. Inner Front Panel
2. Retaining Strip
3. Manual Door Release Interlock Lever