Specifications

J326XP-IM-5AB
4-6
Troubleshooting
Accessing the Rotor in Case of Power Failure
1
Turn the power switch to off (O) and unplug the power cord from the main source.
2
Insert a 4-mm or
5
/32-in. Allen wrench straight through one of the holes (either one) and turn
the wrench to the left (counterclockwise) about one-half turn.
The latch disengages.
3
Repeat step 2, inserting the Allen wrench through the other hole.
After the second latch disengages, the front panel will fall forward about an inch from
the top.
The bottom of the front panel is held in place by three grooved brackets, attached to the
front panel, that seat over a metal retaining strip on the centrifuge bottom panel.
4
Lift the front panel up and off the centrifuge; set it aside.
You will see an inner front panel that extends about halfway down from the centrifuge top
(see Figure 4.2).
Below this inner panel is a black manual door release interlock lever (see Figure 4.2).
Figure 4.2 Manual Door Release
1. Inner Front Panel
2. Retaining Strip
3. Manual Door Release Interlock Lever
1
2
3