Specifications
J326XP-IM-5AB
3-6
Operation
Entering Run Parameters
Timed Run
1
Turn the TIME knob until the required run time appears in the TIME display.
• If you enter more than 59 minutes in the minutes field, the system automatically converts
the entry to hours and minutes after
ENTER is pressed.
2
When all run parameters are entered, press START.
• The rotor begins to accelerate and the display begins showing the remaining run time.
• The run will end when the time counts down to 0 or when STOP is pressed.
Continuous (HOLD) Run
1
Press the TIME/HOLD key to select the HOLD mode.
• The red
HOLD indicator light comes on in the TIME display.
2
When all run parameters are entered, press START.
• The rotor begins to accelerate and the display begins showing the elapsed time.
• The time will count up to 3 hours, 0 minutes, and then will reset to 0 and continue counting.
3
Press STOP to end the run.
• The rotor decelerates to a stop.