Specifications

J326XP-IM-5AB
3-6
Operation
Entering Run Parameters
Timed Run
1
Turn the TIME knob until the required run time appears in the TIME display.
If you enter more than 59 minutes in the minutes field, the system automatically converts
the entry to hours and minutes after
ENTER is pressed.
2
When all run parameters are entered, press START.
The rotor begins to accelerate and the display begins showing the remaining run time.
The run will end when the time counts down to 0 or when STOP is pressed.
Continuous (HOLD) Run
1
Press the TIME/HOLD key to select the HOLD mode.
The red
HOLD indicator light comes on in the TIME display.
2
When all run parameters are entered, press START.
The rotor begins to accelerate and the display begins showing the elapsed time.
The time will count up to 3 hours, 0 minutes, and then will reset to 0 and continue counting.
3
Press STOP to end the run.
The rotor decelerates to a stop.