Specifications
J326XP-IM-5AB
3-5
Operation
Entering Run Parameters
3
3
Turn the SPEED knob until the required RPM appears in the SPEED display.
• When the run is started, the centrifuge operates at the set speed.
• The corresponding RCF is automatically calculated.
a. Press
RPM/RCF to display the RCF value.
Setting RCF
1
Check to see if the red RCF light is lit on the display.
a. If it is lit, go to step 3.
2
Press the RPM/RCF key to select the RCF mode.
• On the display, the red RCF ( g) LED lights.
3
Turn the SPEED knob until the required RCF value appears in the SPEED display.
• The centrifuge operates at a speed calculated to produce the set RCF value.
a. Press
RPM/RCF to display the RPM value.
Setting Run Time
Pressing the TIME/HOLD key toggles the system between HR:MIN mode and HOLD mode. HR:MIN
mode is the default.