Specifications

J326XP-IM-5AB
3-4
Operation
Entering Run Parameters
Entering Run Parameters
Selecting a Rotor
1
Turn the ROTOR knob until the rotor in use appears in the display.
Small green LEDs by each rotor type abbreviation light one at a time, and the rotors of that
type appear consecutively in the display.
Setting Run Speed
Enter a run speed up to 26 000 rpm, or up to the maximum speed of the rotor in use. Or, enter a
relative centrifugal field (RCF) value up to the maximum achievable RCF of the rotor in use.
Setting RPM
1
Check to see if the red RPM light is lit on the display. (RPM is the default speed mode.)
a. If it is lit, go to step 3.
b. If the centrifuge is already in the
RPM mode, skip step 2 and go on to step 3.
2
Press the RPM/RCF key to select the RPM mode.
On the display, the red
RPM LED lights.
ROTOR
JSP
JS
JA
JLA
JV
JE
JCF
RPM