Specifications
J326XP-IM-5AB
3-2
Operation
Summary of Avanti J-26XP Run Procedures
Summary of Avanti J-26XP Run Procedures
For runs at temperatures other than ambient, precool or prewarm the rotor to the required temperature
before the run.
1
Press the POWER switch to on (I). Open the centrifuge door.
2
Install the rotor. Make sure that the rotor load is balanced and that the rotor is securely fastened
to the centrifuge drive hub. Close the centrifuge door.
3
Turn the Rotor knob until the installed rotor name appears in the ROTOR ID display.
4
Turn the SPEED knob until the required run speed (rpm) appears in the SPEED display. Or press
RPM/RCF to select RCF mode and turn the SPEED knob until the required rcf appears in the
SPEED display.
5
Turn the TIME knob until the required run length appears in the TIME display. Or press
TIME/HOLD for continuous operation.
6
Turn the TEMP knob until the required temperature appears in the TEMP°C display.
7
Press ACCEL repeatedly to select MAX (maximum) or SLOW acceleration.
8
Press DECEL repeatedly to select MAX (maximum) or SLOW deceleration, or select OFF
(no brake).
9
Check that all parameters are correct and that the door is closed. Press START.
10
Wait for the set time to count down to zero, or end the run by pressing STOP.
11
When the run is complete, open the door and unload the rotor.