J326XP-IM-5AB
2-4
Description
Controls and Indicators
Figure 2.1 The Power Switch
Control Panel
The control panel (Figure 2.2) is mounted at an angle on the centrifuge top rear for easy visibility
and access. It contains a digital display for actual and setup values, knobs and touch keys for
parameter entry, and additional touch keys for system control.
Figure 2.2 The Control Panel
Digital Display
The display shows current status of the instrument whenever the power is turned on.
• The display shows both actual centrifuge values and user-set parameters. The
SET/ACTUAL key
is used to toggle between the two views.
• During a run, actual values are displayed. If you turn a knob (other than the
TIME knob) during
operation, the display will change to show the set values. If no parameters are entered or
changed for 5 seconds, the display will revert to showing actual values.
• During a timed run, the
TIME display shows the remaining run time in hours and minutes. The
TIME display begins counting down when START is pressed and continues counting down to 0,
when deceleration begins. When the rotor decelerates to 0 rpm,
End flashes in the TIME display.
1. Digital Display
1