Specifications
J326XP-IM-5AB
1-2
Specifications and Preinstallation Requirements
Specifications
Operational Features
Physical Data
Acceleration maximum or slow
Deceleration maximum, slow, or off
a. During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. Refer to
applicable rotor manuals for specific rotor operating range information.
b. To reach temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber
during operation. At low run speeds or low ambient temperatures, the centrifuge may not be able to achieve some higher
temperatures.
Specifications Description
Specifications Description
Door 6.1-cm (2.4-in.) thick structural foam with steel plate
Rotor Chamber Diameter 51.3 cm (20 in.)
Friction Reduction
System (FRS)
190 mm (7.5 in.) Hg
Specification Description
Width 71 cm (28 in.)
Depth • 86 cm (34 in.)
• including air diverter extending from back panel
— 102 cm (40.25 in.)
Height • with door closed
— 86 cm (34 in.)
• to top of control head
— 116 cm (45.5 in.)
• to top of open door
— 149 cm (58.7 in.)
Weight 290 kg (640 lb)
Clearances
(for adequate
ventilation)
• sides
— 7.6 cm (3 in.)
• back (place air diverter against the wall)
— 16 cm (6.25 in.)
Surface Finish polyester control panel with polycarbonate coating overlay; polyurethane
enamel on door and covers; acrylic baking enamel on other surfaces