Operating instructions
Communication protocol Operation modes and their limitations
KNF Flodos CP_SIMDOS02_EN_01_166851 9
NOTICE
It is not allowed that two pumps can send their answers simulta-
neously
The protocol answer can be deactivated (see command SPn).
In this case no ACK or NACK sign is returned. The reply for com-
mands that asked for information is always returned.
8 Operation modes and their limitations
MS==0: Run Mode
DT: unused
DV: unused
RV is forced to be within the interval [RV
min
, RV
max
] by the firm-
ware.
MS==1: Metering by set volume and time
DT: is forced to be within the interval [DT
min
, DT
max
] by the
firmware.
=
=
DV: set by the user (check limits [DV
min
, DV
max
])
RV is forced to be within the interval [RV
min
, RV
max
] by the firm-
ware.
MS==2: Metering by set flow rate and time
DT: set by the user
DV: calculated by firmware
=
RV is forced to be within the interval [RV
min
, RV
max
] by the firm-
ware.
Legend
DT: Dispense time [s]
RV: Flow rate [µl/min]
DV: Metering volume [µl]
SIMDOS 02
RV
min
0.03ml/min
RV
max
20ml/min
DV
min
0.03ml
DV
max
1000l
SIMDOS 10
RV
min
0.1ml/min
RV
max
100ml/min
DV
min
0.1ml
DV
max
1000l