Operating instructions

Universal commands Communication protocol
6 KNF Flodos CP_SIMDOS02_EN_01_166851
4 Universal commands
The address number 99 has a special function. Commands, which
don’t require an answer, can be sent to address 99. The command
will be carried out by all pumps. This enables synchronized func-
tions such as the start of all pumps at once.
5 Interface parameters
The pump is permanently set to the following values:
Name Value
Baud Rate 9600 Baud
Data Bits 8
Parity No
Stop Bit 1
Handshake: There’s no monitoring. The user is responsible that
the pump is not overloaded.
Timing: The typical reaction time of the pump is 2 ms. If the
answer takes longer than 100 ms, there is either a
problem with the communication or the pump is still
busy with a time consuming function e.g. new pump
initializations.