9.1.28 Maintenance position
Move to maintenance position. When this function is activated, the
eccentric of the pump moves to a position where maintenance is
easily possible.
MPn
MP
n (1 digit, right adjusted)
MP1 ⇒ Move to the maintenance position
n Function
0 No movement to maintenance position (Factory default)
1 Movement to maintenance position is triggered
?MP
n (1 digit, right adjusted)
Transmit ?MP
Receive 1 ⇒ Maintenance position reached and active
n
Function
0 Out of maintenance position
1 Maintenance position reached and active
Display on Pump User Interface:
n Function
0 No display
1 Full screen displays "Maintenance"
Value of MP is not saved
Value of MP is not saved
While a maintenance state:
• A stop command switches the motor torqueless.
• A start command starts the pump normally
• A prime command moves the diaphragm in the montage
position (lower dead centre).