Operating instructions

Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 33
9.1.20 Auto-start after power on
Settings of automatic start after power off/switch off.
With active Auto-start the pump will start automatically if one of the
following actions takes place:
The power plug is attached to the pump
The pump is switched on
NOTICE
If any input control signal is configured the pump will stay in
pause state until a start trigger is given.
SAn
SA
n (1 digit, right adjusted)
SA1 activate Auto-start
n Function
0 Auto-start inactive (Factory default)
1 Auto-start active
?SA
n (1 digit, right adjusted)
Transmit ?SA
Receive SA1 Auto-start is active
n
Meaning
0 Auto-start is inactive
1 Auto-start is active
Display on Pump User Interface:
n Display
0 No display
1 "AS" is displayed above unit indicator
Value of SA is saved
Value of SA is saved
--
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: