Operating instructions

Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 31
9.1.18 Characteristic pump profile selection
Selects the active pump profile
CCn
CC
n (1 digit, right adjusted)
CC1 Profile for volatile fluids is active
n Function
0 Standard profile is active
1 Profile for volatile fluids is active
2 Profile for viscous fluids is active
3 Profile for high viscous fluids is active
4 Reserved
?CC
n (1 digit, right adjusted)
Transmit ?CC
Receive CC1 Profile for volatile fluids is active
n
Meaning
0 Standard profile is active
1 Profile for volatile fluids is active
2 Profile for viscous fluids is active
3 Profile for high viscous fluids is active
4 Reserved
Display on Pump User Interface:
n
Display
0
Agent “Standard
1
Agent “
Degassing
2
Agent "Visc100cSt
3
Agent “
Visc500cSt
4 Reserved
Value of CC is saved
Value of CC is saved
--
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: