Operating instructions
Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 29
9.1.16 Customer level measured calibration volume
Customer level calibration based on Run mode flow rate or Dis-
pense mode volume
CFnnnnnnnn
CF
nnnnnnnn (8 digits, right adjusted)
CF 00000100 ⇒ 100 µl or 100 µl/min, it depends on the mode of
operation
n Function
0...99999999
Customer measured flow rate [µl/min]
or dispense volume in [µl]
No parameter read function
-
-
Display on Pump User Interface:
n Display
0…9999
Measured value nnnn in ml
(when in Dispense mode MS=1)
0…9999
Measured value nnnn in ml/min
(when in Run mode MS=0)
Value of CF is not saved
Value of CFis not saved
The parameter CF is a user input value only.
Setting CF initiates the computation of the calibration parameter
CH. See definition of CH.
NOTICE
CH is computed based on CF (see section 9.1.17).
If the value of CF violates - after computation - the range of CH,
CF is not accepted and a NACK sign is returned.
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: