Operating instructions

Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 27
9.1.14 Open collector output function
Defines the function of the open collector output (low active)
RSn
RS
n (1 digit, right adjusted)
RS1 Output on active level, when motor is running
n Function
0 active signal level when Alarm on Error (Factory default)
1 active signal level when Motor is running
2 active signal level when Volume finish
3 active signal pulse: every 1/10 revolution (10 pulses per
revolution)
4 active signal pulse: 1 pulse per 20µl SIMDOS 02
1 pulse per 100µl SIMDOS 10
?RS
n (1 digit, right adjusted)
Transmit ?RS
Receive RS1 Output on active level, when motor is
running
n Meaning
0 active signal level when Alarm on Error (Factory default)
1 active signal level when Motor is running
2 active signal level when Volume finish
3 active signal pulse: every 1/10 revolution (10 pulses per
revolution)
4 active signal pulse: 1 pulse per 20µl SIMDOS 02
1 pulse per 100µl SIMDOS 10
Display on Pump User Interface:
Display does not indicate any output signals.
IF L1 is set to 01 or 06, the selected option is not accessible for L2.
Value of RS is saved
Value of RS is saved
--
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
User Interface behavior:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: