Operating instructions
Commands Communication protocol
24 KNF Flodos CP_SIMDOS02_EN_01_166851
9.1.12 Digital input 1 function
Selects the function of the digital input 1.
L1nn
L1
nn (2 digits, right adjusted)
L101 ⇒ Digital input 1, is configured for a level controlled
Start/stop signal
nn Function
00 Signal: Off (Factory default)
01 level controlled: Start/stop
06 edge controlled: Start/stop
?L1
nn (2 digits, right adjusted)
Transmit ?L1
Receive L101 ⇒ Digital input 1, is configured for level
controlled Start/stop signals
nn Meaning
00 Signal: Off (Factory default)
01 level controlled: Start/stop
06 edge controlled: Start/stop
Display on Pump User Interface:
Display indicates only if digital input is active or not (high and low).
IF L1 is set to 01 or 06, the selected option is not accessible for L2.
Value of L1 is saved
Value of L1 is saved
L1 and L2 cannot be set to 01 or 06 at the same time. Protocol
answer is NACK in case of conflicting settings.
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
User Interface behavior:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: