Operating instructions
Commands Communication protocol
22 KNF Flodos CP_SIMDOS02_EN_01_166851
9.1.10 Analog control signal type selection
Selects the analog signal type
RAn
RA
n (1 digit, right adjusted)
RA2 ⇒ Analog signal 4…20mA
n Function
0 Analog signal 0…10V (Factory default)
1 Analog signal 0…20mA
2 Analog signal 4…20mA
3 Analog signal 0..5V
9 Analog signal OFF
?RA
n (1 digit, right adjusted)
Transmit ?RA
Receive 2 ⇒ Analog signal 4…20mA
n
Meaning
0
Analog signal 0…10V (Factory default)
1
Analog signal
0…20mA
2
Analog signal 4…20mA
3
Analog signal
0..5V
9
Analog signal OFF
Display on Pump User Interface:
Display indicates only if analog input is active or not.
When analog signal is changed from “off” to another value, then:
RA changes accordingly.
Value of RA is saved
Value of RA is saved
--
NOTICE
If pump is not in “run mode” any value except “analogue off” is not
accepted. (NACK will be returned)
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
User Interface behavior:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: