Operating instructions

Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 21
9.1.9 Actual run or dispense volume counter µl
Volume counter of the Dispense mode (and Run mode). Counter
resets to 0 and starts at dispense start or run start
No command string
--
--
--
?TV
nnnnnnnnn (9 digits, right adjusted)
Transmit ?TV
Receive 000010000 10 ml
nnnnnnnnn
Meaning
0…999‘999999
Dispensed volume in µl
Display on Pump User Interface:
n Display
0…999999.9 Dispensed volume since last pump start in ml
> 999999 If Upper limit of TV is reached the counter stops
Value of TV is not saved
Value of TV is not saved
Since the pump only dispenses during pressure strokes, the dis-
pense counter needs calibration. The counter is based on the
calibrated pump stroke volume which is summed up by incremental
steps.
Take note of the upper limit. If the dispensed volume is larger than
the upper limit of TV, the volume counter displays the difference to
1000 l and the display shows “> 1000 l”
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: