Operating instructions

Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 19
9.1.7 Dispense mode: Break time between volumes
Break time between two programmed dispense cycles
DBnnnnn
DB
nnnnn (5 digits, right adjusted)
DB00010 10 seconds break time
nnnnn Function
1...5999
Break time in seconds
?DB
nnnnn (5 digits, right adjusted)
Transmit ?DB
Receive 00010 10 seconds break time
nnnnn
Meaning
1...5999 Break time in seconds
Display on Pump User Interface:
nnnnn Display
1...5999 Break time in seconds, is counting backwards when
active
Value of DB is saved
Value of DB is saved
--
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read Command
Command string:
Answer:
Example:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: