Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 19
9.1.7 Dispense mode: Break time between volumes
Break time between two programmed dispense cycles
DBnnnnn
DB
nnnnn (5 digits, right adjusted)
DB00010 ⇒ 10 seconds break time
nnnnn Function
1...5999
Break time in seconds
?DB
nnnnn (5 digits, right adjusted)
Transmit ?DB
Receive 00010 ⇒ 10 seconds break time
nnnnn
Meaning
1...5999 Break time in seconds
Display on Pump User Interface:
nnnnn Display
1...5999 Break time in seconds, is counting backwards when
active
Value of DB is saved
Value of DB is saved
--
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read Command
Command string:
Answer:
Example:
Power OFF behavior:
Auto-start Power OFF behavior: