Operating instructions

Commands Communication protocol
16 KNF Flodos CP_SIMDOS02_EN_01_166851
9.1.5 Dispense mode time for dispensing a volume
Sets the set value for the time to dispense a volume in Dispense
mode.
DThhmmssss
DT
hhmmssss (8 digits, [hh:mm:ss.ss] “h”: hour, “m”: minute,
“s”:seconds, “.ss”: 1/100 seconds)
DT00010000 1 min
hhmmssss Function
99‘99’99’99 Set value for the time to dispense a volume in
Dispense mode
99’59’59’99 Maximum accepted value (NACK if higher)
00’00’01’00 Minimum accepted value (NACK if lower)
00’00’10’00 Factory preset, Auto-start preset (10 seconds)
?DT
hhmmssss (8 digits, right adjusted)
Transmit ?DT
Receive 00010000 1 min
hhmmssss
Meaning
0…99999999 “h”: hour, “m”: minute, “s”:seconds, “.ss”: 1/100
seconds
Display on Pump User Interface:
nn:nn
Display
mm:ss 00:01 - 59:59
Time: mm [min.] ss [sec.]
(when in Dispense mode
MS
= 1 or 2)
hh:mm 01:00 - 99:59
Time: hh [h.] mm [min.]
(when in Dispense mode
MS
= 1 or 2)
off --:--
Time: - - min - - sec
(when in Run mode
MS
= 0)
User Interface behavior:
If time is dialed from “0” to a time value,
then: MS = 1 or 2, DV = RV
If time is dialed to “0”, then: MS = 0, RV = DV, display time “off”
Value of DT is saved
Value of DT is saved
Summary:
Set command
Command string:
Parameter name:
Parameter value:
Example:
Read command
Command string:
Answer:
Example:
Power OFF behavior:
Auto-start Power OFF behavior: