9.1.4 Dispense mode dispense volume
µ
l
Sets the set value for the dispense volume in Dispense-mode. The
unit is µl.
DVnnnnnnnn
DV
nnnnnnnn (8 digits, right adjusted)
DV00020000 20’000 µl
SIMDOS10
nnnnnnnn Function
0...99’999’999 Set value for the dispense volume in Dispense
model]
00’999’999 Maximum accepted value (NACK if higher)
00’001’000 Minimum accepted value (NACK if lower)
00’010’000 Factory preset, Auto-start preset
SIMDOS02
nnnnnnnn
Function
0...99’999’999 Set value for the dispense volume in Dispense
mode
00’999’999 Maximum accepted value (NACK if higher)
00'000'030 Minimum accepted value (NACK if lower)
00’10’000 Factory preset, Auto-start preset
?DV
nnnnnnnn (8 digits, right adjusted)
Transmit ?DV
Receive 00020000 20’000 µl
nnnnnnnn Meaning
min…99999999
Current set value for dispense volume in µl
Display on Pump User Interface:
nnnnnnnn Massage
min … 999.9 Volume nnn.n ml
(when in Dispense mode MS = 1)
Value of DV is saved
Value of DV is saved
Take notice of the limits! See chapter 8.