Operating instructions

Communication protocol Commands
KNF Flodos CP_SIMDOS02_EN_01_166851 13
9.1.3 Run mode flow rate µl/min.
Sets the set value for the flow rate in Run mode. The unit is µl/min.
RVnnnnnnnn
RV
nnnnnnnn (8 digits, right adjusted)
RV00020000 20’000 µl/min
SIMDOS10
nnnnnnnn Function
0...99’999’999 Set value for the flow rate [µl/min] in Run mode
00’100’000 FEM 1.10 Maximum accepted value (NACK if
higher)
00’001’000 FEM 1.10 Minimum accepted value (NACK if
lower)
00’010’000 Factory preset
SIMDOS02
nnnnnnnn Function
0...99’999’999 Set value for the flow rate [µl/min] in Run mode
00'020'000 FEM 1.02 .18 Maximum accepted value
(NACK if higher)
00'000'030 FEM 1.02 .18 Minimum accepted value
(NACK if lower)
00’10’000 Factory preset
?RV
nnnnnnnn (8 digits, right adjusted)
Transmit ?RV
Receive 00020000
nnnnnnnn Meaning
0…99999999
Current set value for flow rate in µl/min
Summary:
Set command
Command String
Command name:
Parameter value:
Example:
Read Command
Command String
Answer:
Example: