Operating instructions

Commands Communication protocol
12 KNF Flodos CP_SIMDOS02_EN_01_166851
9.1.2 Start, Stop, Prime/Drain
To initiate primary pump function. Start pumping, stop pumping,
start Prime/Drain cycle.
KY
n (1 digit, right adjusted)
KYn
KY2 Priming the pump
n Function
0 Stop (Factory preset)
1 Start (Auto-start preset)
2 Prime/Drain (1 stroke)
3 Pause
No parameter read function
Display on Pump User Interface:
n
Display
0 Volume counter stops and the STOP indicator is
active
1 Volume counter either resets to
0 or keeps its value
and starts counting corresponding to the previous
state of pump (either 0 (STOP) or 3 (PAUSE))
2 Volume counter resets to 0 and finally displays the
dispensed volume of the prime stroke
3 Volume counter xxx stops, is blinking and the
PAUSE indicator is active
State of KY reset to factory preset (Always Stop when power ON).
State of KY reset to factory preset (Always Stop when power ON).
See AS (Section 9.1.20) for more information on autostart
Summary:
Set command
Parameter name:
Parameter value:
Command string:
Example:
Read Command:
Command string:
Power OFF behavior:
Auto-start Power OFF behavior:
Remarks: